SUMMARY FOR ELECTIVE SURGERY CONT...
Reprogram Pacemakers
- asynchronous mode
- disable adaptive rate pacing
- enable magnet responsiveness
Indications for reprogramming
- pacemaker dependent patient
- rate responsive mode
- major chest and abdominal surgery
- special pacing indications ( HOCM,paediatrics)